| Model Number |
LF4200 |
| Medical Device Problem Code |
Energy Output To Patient Tissue Incorrect (1209)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Blood Loss (2597)
|
| Date of Event |
06/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample was requested but to date has not been received for evaluation.If the sample or additional information pertinent to the incident is obtained, a follow-up report will be submitted.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a procedure, the device activated normally but approximately 50cc's of bleeding resulted.Another device of the same type was used to continue the procedure.No serious injury occurred or medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one used lf4200 ligasure device was received, and evaluation of the returned sample could not confirm the reported issue.Visual inspection found no defects.The device was activated multiple times on porcine kidney samples, pressing the button in various locations on a range of vessel sizes to detect any activation issues.All seal cycles were completed satisfactorily, and a visual seal effect with an end tone was observed for each application.The investigation found the device to function normally and within specifications.There are factors during use that can affect seal quality, and the instructions for use included with this product lists measures to ensure sealing effectiveness.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|