• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER LIGASURE IMPACT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER LIGASURE IMPACT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF4200
Medical Device Problem Code Energy Output To Patient Tissue Incorrect (1209)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Blood Loss (2597)
Date of Event 06/26/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample was requested but to date has not been received for evaluation.If the sample or additional information pertinent to the incident is obtained, a follow-up report will be submitted.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer reported that during a procedure, the device activated normally but approximately 50cc's of bleeding resulted.Another device of the same type was used to continue the procedure.No serious injury occurred or medical intervention was required.
 
Additional Manufacturer Narrative
Evaluation summary: one used lf4200 ligasure device was received, and evaluation of the returned sample could not confirm the reported issue.Visual inspection found no defects.The device was activated multiple times on porcine kidney samples, pressing the button in various locations on a range of vessel sizes to detect any activation issues.All seal cycles were completed satisfactorily, and a visual seal effect with an end tone was observed for each application.The investigation found the device to function normally and within specifications.There are factors during use that can affect seal quality, and the instructions for use included with this product lists measures to ensure sealing effectiveness.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE IMPACT
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key6721749
Report Number1717344-2017-05602
Device Sequence Number187461
Product Code GEI
UDI-Device Identifier10884521103351
UDI-Public10884521103351
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K031011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2017
Device Model NumberLF4200
Device Catalogue NumberLF4200
Device Lot Number239458LX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2017
Initial Date Received by Manufacturer 06/26/2017
Supplement Date Received by Manufacturer09/22/2017
Initial Report FDA Received Date07/18/2017
Supplement Report FDA Received Date10/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age57 YR
-
-