• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. IMPRA CARBOFLO EPTFE VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. IMPRA CARBOFLO EPTFE VASCULAR GRAFT Back to Search Results
Model Number F5006TWSC
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during preparation of a surgical graft, a tear in the graft occurred while removing the beading.A new surgical graft was prepped and implanted without incident.There was no reported patient contact.
 
Additional Manufacturer Narrative
Manufacturing review: the lot number has been provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation and two photos were provided for review.One segment of graft was returned with most of the beading returned.A tear was noted along the beading track of the segment.Therefore, the investigation is confirmed for torn material.The definitive root cause could not be determined based upon available information.It is unknown whether procedural issues contributed to the event.Labeling review: the current instructions for use (ifu) states: precautions: when removing the external spiral support (beading) of the graft, the beading must be removed slowly and at a 90° angle to the graft.Rapid unwinding and /or removal at less than a 90° angle may result in graft damage.Do not use surgical blades or sharp, pointed instruments to remove the beading as this may damage the graft wall.If damage occurs, that segment of the graft should not be used.
 
Event or Problem Description
It was reported that during preparation of a surgical graft, a tear in the graft occurred while removing the beading.A new surgical graft was prepped and implanted without incident.There was no reported patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRA CARBOFLO EPTFE VASCULAR GRAFT
Common Device Name
EPTFE VASCULAR GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6722324
Report Number2020394-2017-00811
Device Sequence Number187551
Product Code DSY
UDI-Device Identifier00801741021558
UDI-Public(01)00801741021558
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K004011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/28/2021
Device Model NumberF5006TWSC
Device Catalogue NumberF5006TWSC
Device Lot NumberVTAY0697
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2017
Initial Date Received by Manufacturer 06/20/2017
Supplement Date Received by Manufacturer09/12/2017
Initial Report FDA Received Date07/18/2017
Supplement Report FDA Received Date09/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-