ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT300
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Codes Visual Impairment (2138); No Code Available (3191)
Date of Event 06/06/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (model zct300 +21.5 diopters) was explanted in a secondary surgical procedure as the female patient had poor visual acuity.Reportedly, the surgeon noticed a crack in the edge of the iol.Another lens, same model smaller diopter (+19.0), was implanted as a replacement.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 08/31/2017.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection showed that the lens was cut in pieces, most probably to make the explant possible.Considering the condition of the lens additional analysis was not possible.The customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Additional Manufacturer Narrative
Additional information was received and it was learnt that the intraocular lens (model zct300 +21.5 diopters) was explanted because of unexplained refractive error.There were no complications such as incision enlargement, vitrectomy, or capsule tear when removing the lens.Reportedly, the patient was doing well when discharged.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6723443
Report Number9614546-2017-00631
Device Sequence Number220031
Product Code HQL
UDI-Device Identifier05050474553354
UDI-Public(01)05050474553354(17)180221
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date02/21/2018
Device Model NumberZCT300
Device Catalogue NumberZCT300U215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2017
Initial Date Received by Manufacturer 06/23/2017
Supplement Date Received by Manufacturer10/03/2017
12/05/2017
Initial Report FDA Received Date07/18/2017
Supplement Report FDA Received Date10/26/2017
12/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-