Model Number 520-01-010
Health Effect - Clinical Code No Information (3190)
Date of Event 06/19/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to cuff arthropathy.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient had cuff arthropathy.The in-vivo length of patient service for the implant was 3.6 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to the cuff arthropathy.There are multiple factors that may contribute to the event that are outside the control of djo surgical are patient activities, abnormal glenohumeral wear, humeral head collapse, excessive load or torque, poor bone quality, or trauma.Other conditions, root cause, relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PRIMARY HUMERAL STEM, SZ 10
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6725951
Report Number1644408-2017-00593
Device Sequence Number16403969
Product Code KWS
UDI-Device Identifier00888912024891
UDI-Public(01)00888912024891
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Model Number520-01-010
Device Catalogue Number520-01-010
Device Lot Number455G1096
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/19/2017
Supplement Date Received by Manufacturer08/10/2017
Initial Report FDA Received Date07/19/2017
Supplement Report FDA Received Date09/08/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000,LOT 878C1233; 520-01-238,LOT 190G1045; 520-42-116,LOT 913C1028
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-