| Model Number |
ZCT525 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to the manufacturer has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zct525 13.5 diopter intraocular lens (iol) was explanted on (b)(6) 2017.Reportedly, lens was defective; however, no specifics on the lens was provided.The lens was replaced with a zct400 iol with the same diopter size.There was no incision enlargement, no suture, and no patient injury reported.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation?: yes, returned to manufacturer on 08/01/2017.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye showed the product is cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|