• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G55738
Medical Device Problem Code Wrinkled (2614)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/02/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) # k142688.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195 importer site establishment registration number: 3005580113.1 x echo-hd-25-c was returned to cirl for evaluation.On evaluation of the returned device, it was noted that no syringe was returned with the device.The stylet was partially inserted on return.There was no needle exposure from the sheath.There was a slight resistance when advancing the needle.There was no damage noticed to the outside of the device.There was resistance when re-inserting the stylet.The needle was crumpled- this was the reason for the resistance when re-inserting the stylet.Crumpling was within the handle.The stylet inserts distally with no resistance.The crumpling occurred 5cm from the base of the hub.The customer complaint was confirmed as the handle was crumpled.A definitive root cause for the customer complaint could not be determined as the exact operational conditions of use could not be replicated in the laboratory setting, possible causes may be due to handling of the device or possibly use of excessive force.Prior to distribution, all echo-hd-25-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.The notes section of the instructions for use instructs the user to inspect the device prior to use for any damage: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".From the information provided, no adverse effects to the patient have been reported as a result of this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
The user punctured the pancreas head using echotip procore high definition ultrasound biopsy needle through the descending part of the duodenum.Resistance was met but he stroked several times.Then he detached the device from the endoscope and attempted to insert the stylet to collect the specimens.However the stylet would not advance any further than the middle of the device.He replaced it with another device to complete the procedure.The device was evaluated "on" the 23 jun 2017 and the needle was confirmed to be crumpled in the handle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key6727900
Report Number3001845648-2017-00295
Device Sequence Number11105372
Product Code FCG
UDI-Device Identifier00827002557383
UDI-Public(01)00827002557383(17)190628(10)C1246465
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberG55738
Device Catalogue NumberECHO-HD-25-C
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/20/2017
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2017
Initial Report FDA Received Date07/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-