• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT KYPHON NEUCHATEL MFG KYPHON XPANDER¿ INFLATABLE BONE TAMP; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT KYPHON NEUCHATEL MFG KYPHON XPANDER¿ INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number K13A
Medical Device Problem Code Bent (1059)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that on (b)(6) 2017, patient presented with pre-op diagnosis as: ankylosing vertebral hyperostosis and intravertebral clefts.For which patient underwent balloon kyphoplasty.Intra-op, after balloon was inserted in the vertebral body, the balloon was able to be inflated without problem.When deflating balloon, the marker (made of platinum iridium alloy) looked somewhat distorted.It was removed and checked visually.Although no breakage was observed, the shaft part was bent a little.Repeated removal and insertion was not conducted.The marker before inflation was normal.Product came in contact with the patient and no patient complications were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER¿ INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MDT KYPHON NEUCHATEL MFG
pierre-à-bot 97
neuchâtel,ne 2000
CH  2000
MDR Report Key6728002
Report Number2953769-2017-00080
Device Sequence Number762272
Product Code HRX
UDI-Device Identifier00643169103078
UDI-Public00643169103078
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K041454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberK13A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2017
Initial Report FDA Received Date07/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
-
-