Catalog Number 21-7231-24 |
Device Problem
Material Fragmentation (1261)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 04/01/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
It was reported that the event occurred at the end of (b)(6) 2017.The exact date is unknown.The device is currently being evaluated; smiths medical will file a follow-up report detailing the results of the evaluation once it is completed.
|
|
Event Description
|
It was reported that the catheter of a cleo® 90 infusion set remained in the patient's skin.The device had been in use for "three days or more" when the fault occurred.It was noted that the nurse went to the patient's home later, and the "catheter set" had been done successfully.The cannula worked itself out of the skin.No permanent injury was reported.
|
|
Manufacturer Narrative
|
Three unused cleo® 90 infusion sets were returned for investigation.Examination of the entire devices found no abnormalities or defects.The trocar was in proper position as was the site, soft cannula, etc.Microscopic examination of the soft cannulas were all found to be in proper order with no visual defects or abnormalities.No device issues were found.No root cause for reported problem was determined.
|
|