• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH PATHROMTIN SL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH PATHROMTIN SL Back to Search Results
Model Number PATHROMTIN SL
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The operator contacted a siemens healthcare diagnostics inc.(siemens) technical application specialist (tas) to determine which result they should report for activated partial thromboplastin time (aptt) and whether the pathromtin sl reagent can be used to monitor low-molecular weight (lmw) heparin therapy.The operator indicated that they routinely use the actin fsl reagent to test for aptt and they use the pathromtin sl reagent specifically to test heparinized patients and for this patient they obtained different results when they tested the patient's blood in parallel using the pathromtin sl and actin fsl reagents.The operator did not provide the lot numbers of the pathromtin sl and actin fsl reagents.Siemens has determined that actin fsl and pathromtin sl reagents are not recommended for monitoring of aptt when the patient is on low-molecular weight (lmw) heparin therapy.The instruction for use (ifu) for the actin fsl and pathromtin sl reagents do not have performance claims for monitoring lmw heparin therapy.Siemens offers four different aptt reagents which provide options in sensitivity to heparin, lupus anticoagulants, and factor deficiencies that allows customers to select a reagent type that is most appropriate for monitoring their specific patient population.The variation in sensitivity can lead to differences in the aptt results on patient samples.The instrument and reagents are performing according to specifications.No further evaluation of this device is required.Mdr 9610806-2017-00072 and mdr 9610806-2017-00073 were filed for the same event.
 
Event or Problem Description
A patient was administered low-molecular weight (lmw) heparin therapy after receiving surgery and tested for activated partial thromboplastin time (aptt).In parallel, the operator analyzed the patient blood for aptt using the actin fsl and pathromtin sl reagents and obtained different aptt results.The operator obtained an aptt result of 152 seconds using the pathromtin sl reagent and obtained an aptt result of 30 seconds using the actin fsl reagent.It is unknown if any results were reported to the physician.The operator indicated they explained the situation to the physician such that these aptt results did not affect the patient.The operator indicated that the physician treated the patient according to standard medical protocols and other test results.There are no known reports of patient intervention or adverse health consequences due to the different aptt results obtained between the pathromtin sl and actin fsl reagents.
 
Additional Manufacturer Narrative
Siemens healthcare diagnostics inc.(siemens) filed the initial mdr 9610806-2017-00074 on july 20, 2017.July 21, 2017 additional information: upon further investigation, siemens determined that there was no potential for death or serious injury for this incident.There is negligible potential for injury if this type of incident were to recur.Supplemental mdrs 9610806-2017-00072_s1 and 9610806-2017-00073_s1 were filed for the same event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATHROMTIN SL
Common Device Name
PATHROMTIN SL
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
emil-von-behring- str. 76
marburg, D-350 41
GM  D-35041
MDR Report Key6730214
Report Number9610806-2017-00074
Device Sequence Number670479
Product Code GFO
UDI-Device Identifier00842768007521
UDI-Public00842768007521
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K955450
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 07/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPATHROMTIN SL
Device Catalogue Number10446066
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/21/2017
Supplement Date Received by Manufacturer07/21/2017
Initial Report FDA Received Date07/20/2017
Supplement Report FDA Received Date07/31/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age80 YR
-
-