The reason for this revision surgery was the incorrect poly liner was impacted into the acetabular shell and had to be removed and replaced with the correct liner.The in-vivo length of patient service for the implant was one day or less.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient had an incorrect size insert implanted and the surgeon performed this revision to implant the correct implant.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
|