LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 06/22/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was unable to confirm the reported issue.Functional verification testing was completed without issues.The customer reaffirmed that the set value for co2 was zero.As a precaution.The device was returned to livanova deutschland for further evaluation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that the s5 gas blender system intermittently displayed a co2 alarm during setup.The customer reported that co2 has never been used and restarting the device did not resolve the issue.The device was replaced with a spare.There was no patient involvement.
 
Manufacturer Narrative
The device was returned to livanova (b)(4) for further investigation.During investigation a livanova service engineer was able to confirm the reported issue.As root cause was a defective flow controller identified.Therefore the mass flow controller was replaced to resolve the issue.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6731736
MDR Text Key80809121
Report Number9611109-2017-00555
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/29/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse