Model Number 25-40-45 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/22/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to livanova (b)(4) for further investigation.
During investigation a livanova service engineer was able to confirm the reported issue.
As root cause was a defective flow controller identified.
Therefore the mass flow controller was replaced to resolve the issue.
A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|
|
Manufacturer Narrative
|
There was no patient involvement.
Livanova (b)(4) manufactures the s5 gas blender system.
The incident occurred in (b)(6).
This medwatch report is being filed on behalf of livanova (b)(4).
A livanova field service representative was dispatched to the facility to investigate.
The service representative was unable to confirm the reported issue.
Functional verification testing was completed without issues.
The customer reaffirmed that the set value for co2 was zero.
As a precaution.
The device was returned to livanova deutschland for further evaluation.
If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that the s5 gas blender system intermittently displayed a co2 alarm during setup.
The customer reported that co2 has never been used and restarting the device did not resolve the issue.
The device was replaced with a spare.
There was no patient involvement.
|
|
Search Alerts/Recalls
|