| Model Number |
4134105 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/09/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.A lot history review (lhr) of rexc0823 showed no other similar product complaint(s) from this lot number.Device not returned, at this time.
|
| |
|
Event or Problem Description
|
|
It was reported resistance of the introducer by passing through the skin.By removing it was observed that the tip of the introducer was splited.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a damaged excalibur introducer sheath was confirmed, but the exact cause was unknown.One unsplit excalibur introducer sheath was returned for investigation.The sheath extended 3.1cm from the distal end of the molded green tabs.The needle had been removed from the sheath and was not returned for investigation.The edge at the distal tip of the excalibur introducer sheath was rough and uneven.The sheath was split 1.5mm along one of the score lines at the distal tip.This type of damage can occur if the sheath is advanced over the needle before it enters the vein.The damage most likely occurred during use.Without the needle, it cannot be determined if other factors were involved in the reported event.The product ifu states, ¿grip only on the needle hub during insertion.Do not apply excessive pressure to the wings.¿ a lot history review (lhr) of rexc0823 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported resistance of the introducer by passing through the skin.By removing it was observed that the tip of the introducer was splited.No patient injury was reported.
|
| |
|