Model Number 25-40-45 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/28/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 gas blender system.(b)(4).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The device has been requested for return to livanova (b)(4) for further investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 gas blender system deviated from the desired set point during priming.There was no patient involvement.
|
|
Manufacturer Narrative
|
The device manufacture date was inadvertently left blank in the initial report, submitted july 21, 2017.The device manufacture date is october 11, 2010.
|
|
Manufacturer Narrative
|
The defective gas blender system was returned to livanova (b)(4) for further investigation.During the investigation the reported failure could be confirmed and could be traced to a deviation of the flow controller.The improper calibration of the flow controller was identified as root cause for the reported issue.The device has been disinfected and cleaned and a new adjustment of the flow controller has been performed.Subsequent functional verification testing was completed without further issues and the unit was returned to the customer.
|
|