C.R. BARD, INC. (GFO) LUTONIX 035 DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9004
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Reocclusion (1985); Thrombosis (2100); Claudication (2550)
Date of Event 10/24/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed 3 additional reocclusion complaints were associated with this lot and 1 of the 3 is for the same patient.A review of the device history record (dhr) shows the lot was manufactured to specification.Conclusion: the actual samples were not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was not related to the study device or procedure.Based on the instructions for use (ifu), the occurrence of a reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided will all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, three lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the right superficial femoral artery (sfa).Approximately 27 months after the index procedure, the patient¿s right sfa vessel was reportedly reoccluded.A revascularization was performed and the hcp deemed it was successful.The investigator assessed that the event was not related to the study device or procedure.The sample was discarded by the user facility and is not available for evaluation.No adverse patient effects were reported.This is one of three products involved with the reported event and the associated manufacturers report number are 3006513822-2017-00141 and 3006513822-2017-00143.
 
Additional Manufacturer Narrative
Additional data: relevant tests and lab data has been updated from "unknown" to " angiography on (b)(6) 2016.Angiography on (b)(6) 2017." f.10 (b)(4).Corrected data: date of event has been changed from "(b)(6) 2017" to "(b)(6) 2016." event description was changed from: " it was reported through a clinical registry that during the index procedure, three lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the right superficial femoral artery (sfa).Approximately 27 months after the index procedure, the patient¿s right sfa vessel was reportedly reoccluded.A revascularization was performed and the hcp deemed it was successful.The investigator assessed that the event was not related to the study device or procedure.The sample was discarded by the user facility and is not available for evaluation.No adverse patient effects were reported.This is one of three products involved with the reported event and the associated manufacturers report number are 3006513822-2017-00141 and 3006513822-2017-00143.To "it was reported through a clinical registry that during the index procedure, three lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the right superficial femoral artery (sfa).Approximately 27 months after the index procedure, the patient developed claudication in their right leg.It was discovered the patients right sfa vessel was reportedly reoccluded.A revascularization was performed involving the use of an abbott armada pta balloon and the placement of two stents.The hcp deemed it was successful.The investigator assessed that the event was not related to the study device or procedure.The cec has adjudicated the reocclusion to be possibly related to the study device and procedure.Approximately 4 months after the revascularization, the patients right sfa had a thrombosis.The pi assessed the event was not related to the study device or procedure, but the cec adjudicated the thrombosis to be possibly related to the study device and procedure.Upon further evaluation, it should be noted that the thrombosis occurred after the first revascularization involving adjunctive treatments.Furthermore, one of the adjunctive treatments involved the placement of two stents.The samples were discarded by the user facility and are not available for evaluation.No adverse patient effects were reported.This is one of three products involved with the reported event and the associated manufacturers report number are 3006513822-2017-00141 and 3006513822-2017-00143." conclusion: conclusion was changed to include the following statements: "however, the cec has adjudicated the reocclusion to be possibly related to the study device and procedure" and "a review of the." analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed 3 additional reocclusion complaints were associated with this lot and 1 of the 3 is for the same patient.A review of the device history record (dhr) shows the lot was manufactured to specification.Conclusion: the actual samples were not received for evaluation.A review of the dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was not related to the study device or procedure.However, the cec has adjudicated the reocclusion to be possibly related to the study device and procedure.Based on the instructions for use (ifu), the occurrence of a reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided will all relevant information.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, three lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the right superficial femoral artery (sfa).Approximately 27 months after the index procedure, the patient developed claudication in their right leg.It was discovered the patients right sfa vessel was reportedly reoccluded.A revascularization was performed involving the use of an abbott armada pta balloon and the placement of two stents.The hcp deemed it was successful.The investigator assessed that the event was not related to the study device or procedure.The cec has adjudicated the reocclusion to be possibly related to the study device and procedure.Approximately 4 months after the revascularization, the patients right sfa had a thrombosis.The pi assessed the event was not related to the study device or procedure, but the cec adjudicated the thrombosis to be possibly related to the study device and procedure.Upon further evaluation, it should be noted that the thrombosis occurred after the first revascularization involving adjunctive treatments.Furthermore, one of the adjunctive treatments involved the placement of two stents.The samples were discarded by the user facility and are not available for evaluation.No adverse patient effects were reported.This is one of three products involved with the reported event and the associated manufacturers report number are 3006513822-2017-00141 and 3006513822-2017-00143.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
MDR Report Key6733819
Report Number3006513822-2017-00142
Device Sequence Number1371932
Product Code ONU
UDI-Device Identifier00801741088636
UDI-Public(01)00801741088636
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2015
Device Model Number9004
Device Catalogue Number9020413500080
Device Lot NumberGFXF0886
Was Device Available for Evaluation? No
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer12/13/2017
Initial Report FDA Received Date07/21/2017
Supplement Report FDA Received Date01/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight80
-
-