Catalog Number R5C4479
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was a small bump or roughness located one centimeter left from the lower right corner of a homechoice automated pd set with cassette.This issue was observed before use by the patient and her nurse.The customer was afraid that this irregularity in material could develop into a hole during therapy.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation and an evaluation is complete.A batch review was conducted and there were no deviations found related to the reported condition during the manufacture of this lot.A visual inspection was performed with the naked eye and magnification which revealed a small scratch and indentation in the cassette sheeting.Microscopic inspection showed that this scratch was not a cut.Functional testing was performed which included a simulated therapy.The sample did not leak and therapy was completed without any alarms or issues.The reported condition was verified.An assignable cause of the scratch could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
singapore
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6733940
MDR Text Key80716816
Report Number1416980-2017-06038
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberR5C4479
Device Lot NumberS16L08084
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/16/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-