BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
|
Back to Search Results |
|
| Model Number |
U357592 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Kinked (1339); Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device was returned to the manufacturer but will not be evaluated due to potential cross contamination of an infectious disease from another device.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Sample not evaluated due to potential cross contamination from another device.
|
| |
|
Event or Problem Description
|
|
It was reported that during advancement of the balloon catheter through a 6 fr sheath, strong resistance was met and the catheter allegedly kinked.It was further reported that the health care provider (hcp) attempted to remove the balloon catheter only, but instead removed the catheter, sheath, and guidewire as a single unit, losing access across the lesion.Reportedly, the hcp used another guidewire, sheath, and balloon catheter to complete the procedure.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Visual/microscopic inspection: as the sample was not returned for evaluation, a visual/microscopic inspection could not be performed.Functional/performance evaluation: as the sample was not returned for evaluation, a functional/performance evaluation could not be performed.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the investigation is inconclusive.Due to potential cross contamination with hepatitis c, the returned sample was discarded and not evaluated.The definitive root cause for the identified sheath-related insertion and retraction issue, and kink, could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) states: warnings: do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Precautions: if resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and introducer sheath/guide catheter s a single unit.Use of the ultraverse 035 pta dilatation catheter: position the balloon relative to the lesion to be dilated, ensure the guidewire is in place, and inflate the balloon to the appropriate pressure.Potential adverse reactions: additional intervention.
|
| |
|
Event or Problem Description
|
|
It was reported that during advancement of the balloon catheter through a 6 fr sheath, strong resistance was met and the catheter allegedly kinked.It was further reported that the health care provider (hcp) attempted to remove the balloon catheter only, but instead removed the catheter, sheath, and guidewire as a single unit, losing access across the lesion.Reportedly, the hcp used another guidewire, sheath, and balloon catheter to complete the procedure.There was no reported patient injury.
|
| |
|