| Model Number |
420278-06 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure.The instrument the instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.The clevis did not exhibit any damage or wear marks.Other cables at the wrist were not damaged.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the customer noted a broken cable on the grasping retractor instrument.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
In relation to the reported event, intuitive surgical, inc.(isi) received voluntary report uf (b)(4) on 07/24/2017 with the following event description: while using the robotic grasping forcep, the cable broke at the tip of the instrument.The instrument was and replaced with another forceps.No harm to the patient.There were 5 lives left on the instrument.Procedure - left sided nephrectomy.Device failed (e.G.Broke, couldn't get it to work or stopped working) other than the procedure name there is no additional information provided in voluntary report uf (b)(4) than what was initially reported.Based on this information the complaint will remain reportable due to the following conclusion: the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.This report does not admit that the report or information submitted under this report constitutes an admission that the device, intuitive surgical or intuitive surgical employees, caused or contributed to the reportable event.
|
| |
|
Search Alerts/Recalls
|