COVIDIEN LP LLC NORTH HAVEN V-LOC 90; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number VLOCM1413
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/26/2017
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter: occurred during an otolaryngology surgery.The loop broke.The user tried to use the thread as it was, but the thread broke.The procedure was completed with another device.Patient status is with no problem.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.From the opened product, it was observed, under magnification, that the suture appeared to have split under tension.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LOC 90
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott ave
north haven CT 06473
MDR Report Key6740761
Report Number1219930-2017-06059
Device Sequence Number983342
Product Code GAM
UDI-Device Identifier20884521132785
UDI-Public20884521132785
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100257
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model NumberVLOCM1413
Device Catalogue NumberVLOCM1413
Device Lot NumberA6K0538VFX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2017
Initial Date Received by Manufacturer 06/27/2017
Supplement Date Received by Manufacturer08/10/2017
Initial Report FDA Received Date07/25/2017
Supplement Report FDA Received Date09/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-