MAUDE Adverse Event Report: GETINGE DISINFECTION AB CM320; DISINFECTOR, MEDICAL DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GETINGE DISINFECTION AB CM320; DISINFECTOR, MEDICAL DEVICES Back to Search Results
Model Number CM320
Medical Device Problem Code Component Falling (1105)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being filed under exemption e2016015 by the manufacturer getinge disinfection ab, (registration no.9616031) on behalf of the importer getinge usa, inc., (registration no.3004147784).Additional information will be provided upon results of the investigation.
 
Event or Problem Description
On (b)(4) getinge became aware of an incident caused by one of the medical devices manufactured by getinge.From information provided by the customer after the washing process the wash cart from cm320 washer disinfector had been transferred on the unloading conveyor and moved to the clean side.The wash cart during the proper operation should stopped when reached end of the unloader.However in the case reported by the customer the wash cart overturned and fell down to the ground.Service technician who arrived on site reported that stop pin was broken.As a result of the event no injury has been reported.However we decided to report this case in abundance of caution.
 
Additional Manufacturer Narrative
This report is being filed under exemption e2016015 by the manufacturer getinge disinfection ab, (registration no.9616031) on behalf of the importer getinge usa, inc., (registration no.3004147784).The event is being investigated.Additional information will be provided upon results of the investigation.
 
Additional Manufacturer Narrative
This report is being filed under exemption e2016015 by the manufacturer (b)(4).The event is still being investigated by manufacturing site and currently the issue is consulted with the supplier.Additional information will be provided following the conclusion of the investigation.
 
Additional Manufacturer Narrative
This report is being filed under exemption e2016015 by the manufacturer getinge disinfection ab, (registration no.9616031) on behalf of the importer getinge usa, inc., (registration no.3004147784).Getinge received a customer complaint where, as stated, a cart fell of from the smart trolley.Please note that smart trolley itself is not a registered medical device, however the cm320 device is registered as a medical device system.The smart trolley is an accessory manufactured by an external supplier and labeled by getinge sterilizaton ab and it can be used with a wild range of devices.This issue has been escalated to the trolley's supplier.According to the information provided by the supplier all trolleys are checked after assembly with accordance to the proper protocol.Mentioned protocol included many steps of final inspection which are to be taken before the trolley is released further.It has been established that during final testing general functionality test of the trolley is performed however the retainer ring (circlip) is not checked specifically.It has been also established that there is a possibility that during loading/unloading procedure the circlip could loosen and this may not be noticed by the customer as it is located under the trolley.The circlip is an off-the shelf component.It not taking up any force in this application.Its only purpose is to keep the bolt in place.If the grove is correct and the circlip is correctly installed it simply cannot detach by itself.It is possible that the ring was not properly installed however complaint review did not show a trend of similar complaints.As a result, we cannot exclude ill-judged maintenance actions during the device lifetime.It has been established that when the event took place the device was likely not up to the manufacturer specification and it directly contributed to the outcomes of the event.It has been also established that when the event took place the device was not used for treatment or diagnosis.Given the circumstances and the fact that there is no trend observed for customer complaints with this failure mode, we shall continue to monitor further events of this nature.
 
Additional Manufacturer Narrative
The follow up report (mfr report # 9616031-2017-00026) is submitted due to the missing f/u 3 and it is submitted to correct the error in numbering.There is no new information regarding this case.The final report (3004147784-2017-000266) was sent on 25/06/2018.The date received by mfg is the date when the fda request was received (03.04.2025).
 
Event or Problem Description
Manufacturer's reference number tw: (b)(4).
 
Additional Manufacturer Narrative
This report is being filed under exemption e2016015 by the manufacturer getinge disinfection ab, (registration no.9616031) on behalf of the importer getinge usa, inc., (registration no.3004147784).Root cause analysis of the issue is still under investigation.Additional information will be provided following the conclusion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM320
Common Device Name
DISINFECTOR, MEDICAL DEVICES
Manufacturer (Section D)
GETINGE DISINFECTION AB
ljungadalsgatan 11
vaxjo kronobergs ian [se-07]
vaxjo, 35115
SW  35115
Manufacturer (Section G)
GETINGE DISINFECTION AB
ljungadalsgatan 11
vaxjo kronobergs ian [se-07]
vaxjo, 35115
SW   35115
Manufacturer Contact
dennis genito
45 barbour pond drive
wayne, NJ 07470
9737097515
MDR Report Key6743450
Report Number9616031-2017-00026
Device Sequence Number822332
Product Code MEC
UDI-Device Identifier07340153700215
UDI-Public07340153700215
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Remedial Action Repair
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/25/2018
Report Date (Section F) 07/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberCM320
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/29/2017
Event Location Hospital
Date Report to Manufacturer07/25/2017
Initial Date Received by Manufacturer 06/29/2017
Supplement Date Received by Manufacturer06/29/2017
06/29/2017
04/03/2025
Initial Report FDA Received Date07/26/2017
Supplement Report FDA Received Date09/01/2017
03/19/2018
04/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
Date Report Sent to FDA07/25/2017
-
-