| Model Number |
VP-833 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted with new details if they become available.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter: occurred during an open procedure.Upon first passing through the tissue, the needle detached from the suture.The suture wasn't treated or hydrated prior to use.A new suture was used to complete the procedure with no issue.The patient is alive with no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample noted a suture and one needle.The needle was observed to be detached from the suture.Upon microscopic inspection of the needle, normal needle crimp and swage marks were observed on the needle.It was observed, under magnification, that instrument marks were present on the swaged end of the needle.As no sealed products were returned, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The observed instrumentation damages suggested that the needles were grasped improperly at the swaged end of the needle during application.Firmly grasping the needle at the swaged end can deform the crimp and cause the needle to disengage from the suture.The recommended grasping area of the needle is at the needle body, where the wire is of a full diameter.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|