ENCORE MEDICAL L.P. FOUNDATION HIP; HEAD, FEMORAL, COCR 28/+3.5MM Back to Search Results
Model Number 497-28-350
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/26/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having instability in hip.The surgeon removed and replaced head and liner; a sleeve was added.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient had instability in the hip.The in-vivo length of patient service for the implant was 14.4 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient had instability in the hip.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION HIP
Common Device Name
HEAD, FEMORAL, COCR 28/+3.5MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6744208
Report Number1644408-2017-00616
Device Sequence Number13769999
Product Code LPH
UDI-Device Identifier00888912024044
UDI-Public(01)00888912024044
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2002
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2009
Device Model Number497-28-350
Device Catalogue Number497-28-350
Device Lot Number949821
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/26/2017
Supplement Date Received by Manufacturer08/07/2017
Initial Report FDA Received Date07/26/2017
Supplement Report FDA Received Date09/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
499-28-007,LOT 692531
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-