Exemption number e2016018 b.Braun medical, inc.(importer) is submitting this report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun melsungen ag internal report # (b)(4) we received one used, empty easypump ii st 250-1,5-s without packaging.The provided sample was subjected to a visual examination.Damages that would lead to a malfunction were not detected at the sample.After opening the big white top cap we detected solution at the filling port (lli-cone) and at the patient connector (lla-cone) of the sample.In addition, the sample was taken to a functional test respectively a leak test was carried out.Therefore the pump was filled up to the nominal volume of 250 ml with nacl 0.9 %.After starting the pump, the pump did work immediately (solution was running).Leakages were not detected.Furthermore the flow rate of the pump was tested.Nominal: 175 ml/h actual: 186.1 ml in 1 h the measured value is within our specifications.A too fast flow (as described by the customer) could not be determined at the inspected sample.Device history records (dhr): reviewed device history records, there is no abnormalities and no such defect detected during in process and at final control inspection.
|