Model Number EX1150
Device Problems Failure to Power Up (1476); Sparking (2595)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/16/2017
Event Type  malfunction  
Manufacturer Narrative
Final analysis found burn marks on the ac power adapter which caused the reported inability to power the transmitter.It was concluded that the damage sustained occurred after exposure to a high current flow from the power outlet.
 
Event Description
It was reported that the merlin transmitter was damaged after a lightning strike.After plugging the device to a power source, the whole unit sparked.The transmitter did not power up.The device would be replaced.No further information was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6745419
MDR Text Key81114538
Report Number2017865-2017-07087
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2017
Date Manufacturer Received09/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
-
-