| Model Number |
CC4204A |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Codes |
Corneal Decompensation (1790); Corneal Edema (1791); Blurred Vision (2137); Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
|
| Date of Event |
06/27/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Collamer® ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.(b)(4).Work order search: no similar complaints were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon inserted and removed a cc4204a collamer single piece lens, +23.5 diopter, within the same surgery.The incision was enlarged to remove the lens and a vitrectomy was performed.The lens was not exchanged for another lens.The reporter indicated the lens was torn.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The surgeon reported for the corneal edema there was standard post-operative antibiotic and anti-inflammatory drops used.There was no post-op inflammation.Patient's current condition, as of last visit on (b)(6) 2017, not well; she is 2 months out from the surgery with persistent corneal edema and reduced visual acuity of 20/150.There was no lens or injection system returned.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received - the customer reported the cc4204a collamer single piece lens, was inserted and the lens optic was noted to be broken.Upon removal of the lens, the posterior capsule was torn.There was patient injury - corneal edema, blurry vision and corneal decompensation.An anterior chamber lens was implanted.The customer reported loading the iol caused a fracture of the optic and the cause of the event was the injection system.The patient's current condition is - bcva - 20/150, poor vision, photophobia, foreign body sensation and floaters.Conclusion code: (the collamer single piece iol is generally indicated for primary implantation for the visual correction of aphakia in persons 60 year of age or older in whom a cataractous lens has been removed by cataract extraction.This patient was (b)(6) years old).(b)(4).
|
| |
|