It was reported that the mesh straps came loose from the mesh.The device was not used, another device was used to complete the repair.The sample was returned to davol for evaluation.The pdo ring was not returned with the sample.The straps were attached to the sample with minor fraying noted, there were grasper marks seen on the eptfe.Stitch holes were present as well as a piece of monofilament, evidence of the inner sew line being present.It appears that forces applied during deployment were sufficient to separate the monofilament resulting in material separation of the mesh layers resulting in the separation of the pdo ring from the device.Based on the sample evaluation it appears that when attempting to position the device, forces applied by the user inadvertently caused damaged to the device.This is the only reported complaint for this manufacturing lot of 264 units.
|