AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125300-12
Health Effect - Clinical Code Intimal Dissection (1333)
Date of Event 07/10/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.However, the reported patient effect of dissection is listed in the xience alpine everolimus eluting coronary stent systems instructions for use (ifu) as a known patient effect of coronary stenting procedures.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a mildly calcified, restenosed lesion in the mid left anterior descending (lad) coronary artery.An attempt was made to cross a restenosed lesion with a 3.0x12mm rx xience alpine stent delivery system (sds), but the sds failed to cross due to interaction with the guiding catheter and another previously implanted stent; not due to patient anatomy.The sds was withdrawn without resistance and there was no damage caused to the previously implanted stent.There were no other device issues noted, however, a dissection occurred due to the failure to cross.The lad dissection and stenosis were successfully treated with balloon angioplasty with an unspecified 3.0x12mm balloon followed by placement of an unspecified stent.There was no occurrence of a clinically significant delay.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6745802
Report Number2024168-2017-06226
Device Sequence Number1095655
Product Code NIQ
UDI-Device Identifier08717648199332
UDI-Public08717648199332
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 07/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/06/2019
Device Catalogue Number1125300-12
Device Lot Number6092941
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2017
Initial Report FDA Received Date07/26/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient Weight97
-
-