• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD NOVAFIL; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD NOVAFIL; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number SPB-1233G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/28/2017
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, when the user started suturing, the thread broke.The procedure was completed with another device.There was no patient injury or adverse event.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Under magnification, instrument marks were observed at the swaged end of the needle.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Grasping the needle at the swaged end during application can damage the needle as observed.It is recommended that the needle be grasped from the needle body, where the wire is of a full diameter and significantly stronger than at the swaged end.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVAFIL
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key6746798
Report Number9612501-2017-05863
Device Sequence Number19065424
Product Code GAT
UDI-Device Identifier20884521100777
UDI-Public20884521100777
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K990952
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model NumberSPB-1233G
Device Catalogue NumberSPB-1233G
Device Lot NumberD7C2354X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2017
Initial Date Received by Manufacturer 06/29/2017
Supplement Date Received by Manufacturer09/08/2017
Initial Report FDA Received Date07/27/2017
Supplement Report FDA Received Date10/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-