| Model Number |
SPB-1233G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, when the user started suturing, the thread broke.The procedure was completed with another device.There was no patient injury or adverse event.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Under magnification, instrument marks were observed at the swaged end of the needle.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Grasping the needle at the swaged end during application can damage the needle as observed.It is recommended that the needle be grasped from the needle body, where the wire is of a full diameter and significantly stronger than at the swaged end.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|