One heart reference sensor (hrs) was received for product evaluation.The suspect hrs unit was connected to a known good working clearsight system for analysis and testing.The sensor did not respond appropriately to differences in bladder and sensor height changes.The suspect hrs was flexed near the hdmi connector, thus flexing the internal flat flex cable, and the zero offset values fluctuated.The sensor was able to be zeroed and there were normal blood pressure readings that were obtained.A visual inspection was performed and there was no physical damage and no oil leakage found.An x-ray was taken and there were damaged trace wires that were found on the flat flex cable.The hrs unit failed to function properly prior to being successfully zeroed.The operator¿s manual states that ¿the ev1000 clinical platform ni is intended for use only as an adjunct in patient assessment.This instrument should be used in conjunction with a bedside physiologic monitor.¿ if pressure measurements are inaccurate, the clinician will detect the difference and perform troubleshooting procedures.The root cause of the damaged flat flex cable is repetitive movement or flexing during use.The repetitive movement leads to fractures.A more durable discrete wire design was made and sold as of february 2017.The device service history record review was completed and all manufacturing inspections passed with no non-conformances.The reported issue was confirmed by evaluation.The issue has been escalated and corrective action implemented.This issue will continue to be monitored.
|