| Model Number |
840 |
| Medical Device Problem Code |
Occlusion Within Device (1423)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that, an 840 vent had a severe occlusion alarm.The ventilator was not in use on a patient at the time the event occurred.The covidien service engineer (se) verified the issue.The se inspected the device and replaced the analog interface (ai) printed circuit board (pcb).The unit passed all testing and operated within the manufactures specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation: the analog interface (ai) printed circuit board (pcb) was returned to medtronic¿s product analysis laboratory.A visual inspection was performed and no anomalies were observed.An investigation was performed and the reported issue was not duplicated.No errors were recorded in the diagnostic logs during testing.No fault was found with the returned ai pcb.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|