MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CS INS SIZE 5 13MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CS INS SIZE 5 13MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5531P513
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Injury (2348)
Date of Event 06/28/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The following devices were also listed in this report: tri ts baseplate size 5; cat# 5521-b-500; lot# wuhha, triathlon cr fem comp #4 r-cem; cat# 5510f402; lot# bjt6r, triathlon sym patella s36x10; cat# 5550l360; lot# ley035.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other similar events for the lot referenced.There have been no other events for the sterile lot referenced.It was noted that the device is not available for evaluation.Additional information has been requested and if received, will be provided in the supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that a patient had a revision on a right total knee due to infection.The site was washed out and the poly was replaced.It was reported that there were no surgical delays.
 
Additional Manufacturer Narrative
An event regarding infection involving a triathlon insert was reported.The event was not confirmed.Device evaluation and results: not performed as the device was not returned.Medical records received and evaluation: not performed as no medical records were provided.Device history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the reported lot or sterile lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as x-rays, medical records, operative notes, patient history and pathology reports are required to complete the investigation for determining root cause.A capa trend analysis was conducted for the reported failure mode and concluded infection is a known possible adverse outcome of surgery and is beyond stryker's control.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this investigation will be reopened.
 
Event or Problem Description
It was reported that a patient had a revision on a right total knee due to infection.The site was washed out and the poly was replaced.It was reported that there were no surgical delays.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CS INS SIZE 5 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key6748011
Report Number0002249697-2017-02313
Device Sequence Number10817932
Product Code MBH
UDI-Device Identifier07613327042320
UDI-Public07613327042320
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/25/2021
Device Catalogue Number5531P513
Device Lot NumberLEX819
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/29/2017
Supplement Date Received by Manufacturer08/18/2017
Initial Report FDA Received Date07/27/2017
Supplement Report FDA Received Date09/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
-
-