| Model Number |
33151 |
| Medical Device Problem Code |
Material Erosion (1214)
|
| Health Effect - Clinical Code |
Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
06/14/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The instrument has not been returned for evaluation yet, but the doctor described gross breeches in the insulation of the handle.The device was not inspected before the procedure for damage.We think the insulation damage resulted in arcing and caused burn.
|
| |
|
Event or Problem Description
|
|
Allegedly, the doctor was performing a hysterectomy on a patient when the 33151 handle was laid down on the patient and the cautery activated.There were breeches in the insulation on the handle and it arced and the patient received a burn in the abdomen area.The doctor covered the burn in flamazine and had a cosmetic surgeon assess the burn; the cosmetic surgeon said it would heal nicely and no actions need be taken.Dr.(b)(6) saw her again and says the burn is healing well.He characterized it as a 2nd degree burn.
|
| |
|