MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; HEAD, HUMERAL, OFFSET, 42-20

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; HEAD, HUMERAL, OFFSET, 42-20 Back to Search Results
Model Number 520-42-120
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/28/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient's sub scap failing; the surgeon needed to convert the turon to a reverse total shoulder.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient's subscap failed.The in-vivo length of patient service for the implant was 11 months.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to subscap failure.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors which may cause subscap failure that are outside the control of djo surgical are patient activities, excessive range-of-motion, soft tissue impingement, bone deterioration, trauma or poor bone quality.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
HEAD, HUMERAL, OFFSET, 42-20
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6750765
Report Number1644408-2017-00641
Device Sequence Number1018140
Product Code KWS
UDI-Device Identifier00888912145138
UDI-Public(01)00888912145138
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model Number520-42-120
Device Catalogue Number520-42-120
Device Lot Number915C1053
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/29/2017
Supplement Date Received by Manufacturer08/02/2017
Initial Report FDA Received Date07/28/2017
Supplement Report FDA Received Date09/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000,LOT 878C1599; 520-01-010,LOT 455G1206; 521-01-238,LOT 798N1006
Outcome Attributed to Adverse Event Required Intervention; Disability;
Patient Age66 YR
-
-