| Model Number |
014796 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the procedure the tip of the guide bent.The physician tried to straighten the pointed end but broke it.All pieces were removed from the patient.A backup device was used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to no product being returned the complaint could not be confirmed.Evaluation and investigation are not possible.No definitive conclusions can be made without pertinent manufacturing information or a device to evaluate.No further actions can be taken at this time.A review of the device history record shows that this device passed all acceptance criteria and was compliant upon release for distribution.There are no indications to suggest the device did not meet product specifications nor does it allege any product deficiencies upon release into distribution.
|
| |
|
Additional Manufacturer Narrative
|
|
One (b)(4) 6mm offset endofemoral aimer returned.This is a reusable instrument that is three years old.There was only one portion received.The customer¿s photo shows the broken tip portion.The shaft portion returned has a piece of surgical tape attached where the tip may have been, but is it not with the shaft any longer.The u slide channels have dents and material displaced from overtightening of the universal handle with the aimer.The tip break site is 0.065¿ thick.This indicates a moderate amount of force was used to bend/break this area.Excessive force can result in instrument failure.No root cause related to the manufacture of the device was established.Further investigation is not required at this time.
|
| |
|
Search Alerts/Recalls
|