CORDIS CORPORATION POWERFLEXPRO 8MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4400804S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/04/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that an 8x40mm powerflex balloon catheter ruptured at six atmospheres (6 atm).Therefore the balloon catheter was replaced with a new power flex pro (8x20) and inflated at ten (10) atm.The procedure finished successfully.There was no reported patient injury.The target lesion was unknown.The patient¿s vessel level of tortuousness and calcification was unknown.The rate of stenosis was unknown.A 6f brite tip sheath introducer was inserted and a sv guidewire crossed the lesion.Then under ivus (volcano) the lesion was checked and an 8x100 smart stent was deployed in the target lesion.A power flex pro was delivered to the lesion for post dilatation, however it ruptured.The device prep normally.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.The product was removed intact (in one piece) from the patient.  the product was not returned for analysis.A device history record (dhr) review of lot 17108228 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause of the burst could not be determined.Based on the limited information available for review, vessel characteristics (although unknown) may have contributed to the reported event as calcification/resistance lesions may cause damage to a balloon.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective/preventive action will be taken at this time.
 
Event or Problem Description
It was reported that a power flex balloon catheter ruptured at six atmosphere (6 atm).Therefore the balloon catheter was replaced with a new power flex pro (8*2) and inflated at ten (10) atm.The procedure finished successfully.There was no reported patient injury.A 6f brite tip sheath introducer was inserted and a sv guidewire crossed the lesion.Then under ivus (volcano) the lesion was checked and an 8*10 smart stent was deployed in the target lesion.A power flex pro was delivered to the lesion for post dilatation, however it ruptured.The target lesion was unknown.The patient¿s vessel level of tortuousness and calcification was unknown.The rate of stenosis was unknown.The device prep normally.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient. the product was removed intact (in one piece) from the patient. the product was clinically used and will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 8MM4CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014
MDR Report Key6751547
Report Number9616099-2017-01275
Device Sequence Number10968162
Product Code LIT
UDI-Device Identifier20705032059814
UDI-Public20705032059814
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 07/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2017
Device Model Number4400804S
Device Catalogue Number4400804S
Device Lot Number17108228
Was Device Available for Evaluation? No
Distributor Facility Aware Date07/04/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2017
Initial Report FDA Received Date07/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
(6F BRITE TIP SHEATH) GUIDEWIRE (SV WIRE) IVUS
-
-