• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES HAGENDORF STARDRIVE SCREWDRIVER SHAFT W/HOLDING SLEEVE/T4/66MM; SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES HAGENDORF STARDRIVE SCREWDRIVER SHAFT W/HOLDING SLEEVE/T4/66MM; SCREWDRIVERS Back to Search Results
Catalog Number 03.114.009
Medical Device Problem Codes Dull, Blunt (2407); Device Operates Differently Than Expected (2913); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reportedly there was no patient involvement.Device is an instrument and is not implanted/explanted.Contact phone number: (b)(6).Manufacturing location: (b)(4).Manufacturing date: june 03, 2015.No non-conformance reports were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Lot number 9458292 indicates component 60061590 which is part of the complete part as described in the complaint.A product investigation was completed: a visual inspection under 5x magnification, functional test, drawing review, and dimensional analysis were performed as part of this investigation.A definitive root cause cannot be identified with the limited complaint details.No new, unique or different patient harms were identified as a result of this evaluation.The design is adequate for its intended use when used and maintained as recommended, it did not contribute to the complaint condition.The screwdriver was returned with the distal tip twisted as a result of a clockwise (insertion) fore.It is unlikely that the design contributed to the complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that parts were detected as damaged during internal inspection of loaner kits.Reportedly there was no patient or procedure involvement.This report is for a screwdriver which reportedly was damaged.When the device was received, it was noted that the distal stra drive tip was twisted and stripped.This report is for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARDRIVE SCREWDRIVER SHAFT W/HOLDING SLEEVE/T4/66MM
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH461 4
SZ  CH4614
MDR Report Key6751795
Report Number3003875359-2017-10368
Device Sequence Number1095682
Product Code HXX
UDI-Device Identifier10886982073037
UDI-Public(01)10886982073037(10)9458292
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 10/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.114.009
Device Lot Number9458292
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2017
Initial Report FDA Received Date07/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-