It was reported that the asymptomatic patient presented for a three month implant follow-up.It was noted that the left ventricular lead exhibited noise in the right ventricular channel resulting in the recording of inappropriate high ventricular rate episodes.During initial implant of the pulse generator, the right ventricular and left ventricular leads were reversed and connected into the incorrect ports.The noise was not reproducible in the clinic with isometric and pocket manipulation testing.The clinician reprogrammed the device.On (b)(6) 2017 the patient came in for an unrelated surgery.Upon interrogation, the left ventricular lead was still exhibiting noise, resulting in high ventricular rate episodes.The noise was not reproducible in clinic with isometric and pocket manipulation testing.All other electrical measurements were normal.On (b)(6) 2017 the device was reprogrammed to resolve the issue.
|