| Model Number |
MMT-723RNAP |
| Health Effect - Clinical Code |
Hypoglycemia (1912)
|
| Date of Event |
07/05/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump alarmed button error.The customer¿s blood glucose was 44 mg/dl at the time of incident.Customer stated that the insulin pump alarmed button error and the act button was unresponsive.No significant event leading to button error or keypad anomaly were observed.Button error troubleshooting was performed and confirmed that time is advancing.Customer also reported to have experienced a low blood glucose of 44 mg/dl and treated with food.No low blood glucose troubleshooting was performed due to keypad anomaly.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Unit received no act button response due to corroded keypad traces.No button error alarm.Unit received with minor scratched display window, partially broken reservoir tube lip, cracked reservoir tube, stained end cap sticker and stained address/serial number label.
|
| |
|