| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a call service alarm (call service alarm issue) issue.It was reported that there was a 069 call service alarm.There was no indication that this issue lead to an adverse event.This is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 09/18/2017 with the following findings: a review of the black box history showed multiple call service 087 alarms.The issue was duplicated during power up.The call service 069 failure was written to the black box as a call service 087 failure.Investigation revealed that a language corruption occurred at a component on the printed circuit board resulting in a call service alarm.Unrelated to the original complaint, the battery compartment was cracked below the grip pad.
|
| |
|
Search Alerts/Recalls
|