| Model Number |
2055 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Device Stops Intermittently (1599)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned cable was evaluated.External visual inspection showed no physical damage.The cable failed continuity testing, due to an open in the cable on the yellow conductor inside the bend relief area of the mini-d connector.The cable was connected to a monitor and a "no cable connected" error message was displayed.
|
| |
|
Event or Problem Description
|
|
The customer reported over the past 2 months, they have been acquiring more and more failed cable parts - either through outright cable failures as presented on the zoll unit itself, to inconsistent readings to no readings what-so-ever.No patient impact or consequences were reported.
|
| |
|