• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION POSTERIOR STABILIZED KNEE; INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE6 9MM CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION POSTERIOR STABILIZED KNEE; INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE6 9MM CM Back to Search Results
Model Number 360-09-506
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/03/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the post breaking off from the tibial insert.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the post broke off from tibal insert.The in-vivo length of patient service for the implant was 8 years.The healthcare professional indicated there was a serious risk to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient had broken a ps tibia insert.The surgeon reported no issues associated with the explanted product.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions, root cause, relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION POSTERIOR STABILIZED KNEE
Common Device Name
INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE6 9MM CM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6754166
Report Number1644408-2017-00645
Device Sequence Number10795545
Product Code JWH
UDI-Device Identifier00888912116268
UDI-Public(01)00888912116268
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K933539
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/20/2014
Device Model Number360-09-506
Device Catalogue Number360-09-506
Device Lot Number53980700
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/03/2017
Supplement Date Received by Manufacturer08/30/2017
Initial Report FDA Received Date07/31/2017
Supplement Report FDA Received Date09/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age74 YR
Patient Weight104
-
-