• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION TOTAL SHOULDER; STEM, HUMERAL, PLASMA W/MALE TAPER, 6

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION TOTAL SHOULDER; STEM, HUMERAL, PLASMA W/MALE TAPER, 6 Back to Search Results
Model Number 503-00-006
Health Effect - Clinical Codes Failure of Implant (1924); No Code Available (3191)
Date of Event 07/03/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to a failed total shoulder having been converted to a reverse.
 
Additional Manufacturer Narrative
The reason for this revision surgery was a failed total shoulder converted to reverse.The in-vivo length of patient service for the implant was 8.6 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to a failed total shoulder.There are multiple factors which may cause shoulder failure that are outside the control of djo surgical are patient activities degenerative bone, excessive range-of-motion, low bone density or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to shoulder failure and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION TOTAL SHOULDER
Common Device Name
STEM, HUMERAL, PLASMA W/MALE TAPER, 6
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6754269
Report Number1644408-2017-00646
Device Sequence Number162655
Product Code KWS
UDI-Device Identifier00888912143417
UDI-Public(01)00888912143417
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K950651
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2012
Device Model Number503-00-006
Device Catalogue Number503-00-006
Device Lot Number53830145
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/03/2017
Supplement Date Received by Manufacturer08/21/2017
Initial Report FDA Received Date07/31/2017
Supplement Report FDA Received Date09/19/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
500-04-138, LOT 53938638; 503-38-022, LOT 53946910
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age61 YR
-
-