| Model Number |
RT330 |
| Medical Device Problem Code |
Occlusion Within Device (1423)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/15/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the complaint rt330 optiflow junior tubing kit was returned to fisher & paykel healthcare in (b)(4) for investigation.The complaint rt330 optiflow junior tubing kit was visually inspected.Results: visual inspection revealed that the patient end connector was occluded.Conclusion: the patient end connector was occluded due to a moulding defect.Our investigation has revealed that the patient end connector is a supplied part.Our supplier quality team have been informed of the details of this investigation and will carry out an investigation with the supplier.We have only received one complaint of this nature.The user instructions that accompany the rt330 optiflow junior tubing kit state the following: - check all connections, caps and/or plugs are tight before use - patient monitoring is recommended.
|
| |
|
Event or Problem Description
|
|
A hospital in (b)(6) reported to a fisher & paykel healthcare (fph) field representative that the patient end connector of the rt330 optiflow junior tubing kit had an occlusion.There was no patient involvement.
|
| |
|