Catalog Number 60-6085-202
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device was not returned to conmed for evaluation as it was discarded by the user facility.This complaint is confirmed, however, on the basis of the submitted photo.The device components have clearly detached from the device itself.A review of the manufacturing documents from the device history record has verified the devices were produced according to current and approved procedures and material specifications.Non-conformances regarding the product's identity, quality, safety, effectiveness or performance were not identified during manufacture.There have been no additional complaints reported against this lot of (b)(4) units.A two-year review of complaint history revealed 28 similar complaints have been reported for the reported issue of "cup detachment".(b)(4).A risk analysis was performed and it was found to be acceptable.The instructions for use advise the user of the following.-prior to removal of the device, ensure the locking mechanism is released via the thumbscrew and swipe a finger around the edge of the vaginal cup to separate the tissue from the cup to prevent tissue damage.-do not use excessive force upon device removal to avoid traumatizing the vaginal canal and/or component detachment.-vaginal delivery of a large uterus may result in patient injury.Methods should be used to reduce the size of the uterus prior to removal through the vaginal canal.-upon removing vcare, the surgeon should visually inspect the vcare device, and the patient, to make sure that the entire vcare device was properly removed and that no components or fragments of these components were retained in the patient.There are 5 parts/components to the vcare cervical elevator retractor.The balloon.The forward "cervical" cup, the back or vaginal cup, the locking assembly with thumbscrew, the metal shaft and handle with balloon inflation valve.An investigation is currently opened to determine the root cause of this reported issue.All complaints associated with this product will continue to be monitored through the complaint system.
 
Event or Problem Description
A user facility reported a 60-6085-202 vcare device dropped to bits.The device components were reported to detach from one another but the uterus and upper cup stayed connected.Despite several attempts to collect additional information regarding the term "dropped to bits", no clarification has been provided.The procedure was completed as planned.No patient injury was reported and all detached components were retrieved.This device was discarded by the user facility; therefore, a specific failure mode/root cause is unable to be determined.Although the complaint is unable to be verified/confirmed, photos show the device in multiple pieces.The cervical cup and vaginal cone has completely detached from the manipulator tube.The report is raised on the basis of a device malfunction with potential for injury if recurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCARE LARGE (37MM) CUP
Common Device Name
VCARE
Manufacturer (Section D)
CONMED CORPORATION
525 french road
utica NY 13502 5994
MDR Report Key6758142
Report Number1320894-2017-00169
Device Sequence Number16631256
Product Code LKF
UDI-Device Identifier00653405061819
UDI-Public(01)00653405061819(17)181128(30)1(10)201611281
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K071907
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/28/2018
Device Catalogue Number60-6085-202
Device Lot Number201611281
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2017
Initial Report FDA Received Date08/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age44 YR
-
-