• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED Back to Search Results
Model Number 123622A
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that, upon insertion of the foley catheter, no urine would drain.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "[directions for use]: method of use: the device is intended for single use only and is not reusable.Cleanse the area around the external urethral meatus with the cotton balls immersed in the antiseptics.Lubricate the distal end of the catheter with water-soluble lubricant.Insert catheter into the urethral meatus and advance it until the balloon enters the bladder and urine flows out through the catheter.Using a needle-less syringe, infuse the specified volume of sterile water into the inflation lumen to inflate the balloon.Pull catheter to seat the balloon at the level of the bladder neck and secure placement.To deflate balloon and remove catheter, insert a luer tip (needleless) syringe in the inflation valve to allow the water to drain spontaneously.After balloon deflation, withdraw the catheter while confirming that no resistance is encountered.Precautions for use: when resistance is encountered in inserting catheter, stop the procedure and remove the catheter.When deflating balloon, do not aspirate with a syringe.[the inflation lumen for balloon deflation may be occluded by negative pressure, and as a result the catheter cannot be removed.] when inserting catheter with a stylet, confirm that the stylet has reached the tip of the catheter, and make sure to keep the stylet in place inside the catheter during insertion.No substance except sterile water should be used to inflate the balloon.Do not wipe catheter surface with organic solvents such as alcohol.Do not aspirate urine through drainage funnel wall.Since movement of the body, etc.May twist or bend catheter to cause occlusion, care should be taken to fix the catheter securely.When urinary flow cannot be noted, confirm that the catheter is neither occluded nor broken.When irrigating, open the cap of irrigation funnel and attach irrigation line or tube under aseptic technique.After use, close the cap." (b)(4).
 
Event or Problem Description
It was reported that, upon insertion of the foley catheter, no urine would drain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® FOLEY CATHETER SILICONE COATED
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
lot 57c
covington GA 30014
MDR Report Key6758500
Report Number1018233-2017-04078
Device Sequence Number13956635
Product Code KOD
UDI-Device Identifier00801741095221
UDI-Public(01)00801741095221
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number123622A
Device Catalogue Number123622A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/05/2017
Supplement Date Received by Manufacturer12/27/2017
Initial Report FDA Received Date08/01/2017
Supplement Report FDA Received Date12/29/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
-
-