| Model Number |
840 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that, an 840 ventilator had a blank the upper display.The ventilator was not in use on a patient at the time the event occurred.The service engineer (se) inspected the device and replaced the power supply to correct the issue.The unit passed all testing and operated within the manufacturer specifications.
|
| |
|