ABBOTT MEDICAL OPTICS TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXR00
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implant and explant date: if implanted or explanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was inserted and removed after noting a scratch on it.No further incident.The iol was fully inserted.No incision enlargement, no serious patient injury, no sutures, no vitrectomy.Procedure completed with a lens of the same model and diopter.No additional information was provided to abbott medical optics.
 
Additional Manufacturer Narrative
Device available for evaluation? yes, returned to manufacturer on 08/07/2017.Device returned to manufacturer? yes.Device evaluation: visual inspection with the unaided eye shows the product is cut in half, most probably to make removal and replacement possible.Considering the condition of the lens, additional analysis was not possible.The reported complaint was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The dfu provide the customer with proper usage instructions and guidelines.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6759064
Report Number9614546-2017-00690
Device Sequence Number13825780
Product Code POE
UDI-Device Identifier05050474591684
UDI-Public(01)05050474591684(17)220403
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/03/2022
Device Model NumberZXR00
Device Catalogue NumberZXR00U0120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2017
Initial Date Received by Manufacturer 07/07/2017
Supplement Date Received by Manufacturer09/26/2017
Initial Report FDA Received Date08/01/2017
Supplement Report FDA Received Date10/19/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-