| Model Number |
7500 |
| Health Effect - Clinical Codes |
Abdominal Pain (1685); Nausea (1970); Vomiting (2144); Constipation (3274)
|
| Date of Event |
07/05/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Balloon burst while implanted in patient for 27 days.Product investigation on-going.No serious injury associated with event.Root cause of balloon burst/deflation remains unknown.Possibly related to use error/overpressurization of balloon at placement.Inflation pressures at placement not documented by physician staff.Initial testing suggests that peak pressures balloon exposed to where significantly higher than expected with a single inflation can.Second balloon in placement had an implantation time of 9 days without issue and was found fully inflated at removal.
|
| |
|
Event or Problem Description
|
|
Patient had new onset of non-serious abdominal pain, nausea, vomiting and constipation lasting 4 days.After holiday patient contacted site.X-ray was conducted and one of two balloons was deflated due to a burst.Both balloons were removed as patient wanted to start a new series of balloons.
|
| |
|