ABBOTT MEDICAL OPTICS TECNIS CL; MONOFOCAL IOLS Back to Search Results
Model Number Z9002
Medical Device Problem Code Bent (1059)
Health Effect - Clinical Codes No Patient Involvement (2645); No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) had a bent haptic.No further information was provided.
 
Additional Manufacturer Narrative
Additional information/corrected data: additional information was received and it was learnt there was no patient contact and no patient injury.The haptic was bent before insertion into the eye.This event has now been determined not to be a reportable event.(b)(4).Device available for evaluation? yes.Returned to manufacturer on: 08/03/2017.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection showed that the lens had viscoelastic residue.Slight marks were observed.One haptic was observed bent and the other was observed positioned.The customer's reported complaint was verified however based on the condition of the sample, the issue could not be related to the manufacturing process.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency.No further reports will be submitted for this event.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6759348
Report Number2648035-2017-01417
Device Sequence Number19217392
Product Code HQL
UDI-Device Identifier05050474530546
UDI-Public(01)05050474530546(17)211115
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date11/15/2021
Device Model NumberZ9002
Device Catalogue NumberZ900200125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2017
Initial Date Received by Manufacturer 07/07/2017
Supplement Date Received by Manufacturer08/01/2017
Initial Report FDA Received Date08/02/2017
Supplement Report FDA Received Date08/24/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-