| Catalog Number |
6720-0837 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/05/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a primary implant of the accolade ii 132 angle stem, both outer and inner packaging were found to be compromised.Surgeon used a 127 angle stem instead.Surgical delay was reported to be under 5 minutes with no adverse affects to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding a packaging issue with a accolade stem was reported.The event was confirmed.Device evaluation and results: the device was visually inspected on return.Both the inner and outer tyvek lid were found to be peeled back and open as reported in the event description.Medical records received and evaluation: no medical records were received for review with a clinical consultant.No adverse consequences to the patient were noted.Device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for the lot referenced.Conclusions: nc has been raised to investigate the event.Capa investigation was completed and the following root causes were determined: the primary root-cause for occurrence is "inadequate process and controls when required to inspect, identify, disposition, rework, track and reconcile product that is returned from an fg location".The primary root-cause for escape is "no requirement to have an active halt on lots which are returned from finished goods, or which are transferred into the prms nc warehouse".
|
| |
|
Event or Problem Description
|
|
It was reported that during a primary implant of the accolade ii 132 angle stem, both outer and inner packaging were found to be compromised.Surgeon used a 127 angle stem instead.Surgical delay was reported to be under 5 minutes with no adverse affects to the patient.
|
| |
|