| Model Number |
M00558250 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) investigation result: visual evaluation revealed that the distal end of the needle was slightly bent.Functional analysis revealed that the stylet could be removed and inserted without any issues.The device was able to create and hold vacuum.No sample residue was left in the device.The complaint was not confirmed.Due to anatomical/procedural factors encountered during the procedure; performance of the device was limited.Therefore, the most probable root cause classification is operational context.A review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database confirmed that no other complaints exist for the specified batch.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an expect pulmonary endobronchial ultrasound transbronchial aspiration needle was used in the bronchial tube during an endobronchial ultrasound - transbronchial needle aspiration (ebus-tbna) procedure performed on (b)(6) 2017.According to the complainant, during the procedure there was difficulty puncturing the tissue and was unable to obtain a sample.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient had no complication at the conclusion of the procedure.The investigation found the distal end of the needle was bent, this is now deemed an mdr reportable event.
|
| |
|
Search Alerts/Recalls
|